Evaluation of Innotrac Aio! Second-Generation Cardiac Troponin I Assay: The Main Characteristics for Routine Clinical Use

Author:

Hedberg P.,Valkama J.,Suvanto E.,Pikkujämsä S.,Ylitalo K.,Alasaarela E.,Puukka M.

Abstract

The availability of a simple, sensitive, and rapid test using whole blood to facilitate processing and to reduce the turnaround time could improve the management of patients presenting with chest pain. The aim of this study was an evaluation of the Innotrac Aio! second-generation cardiac troponin I (cTnI) assay. The Innotrac Aio! second-generation cTnI assay was compared with the Abbott AxSYM first-generation cTnI, Beckman Access AccuTnI, and Innotrac Aio! first-generation cTnI assays. We studied serum samples from 15 patients with positive rheumatoid factor but with no indication of myocardial infarction (MI). Additionally, the stability of the sample with different matrices and the influence of hemodialysis on the cTnI concentration were evaluated. Within-assay CVs were 3.2%–10.9%, and between-assay precision ranged from 4.0% to 17.2% for cTnI. The functional sensitivity(CV=20%)and the concentration giving CV of 10% were approximated to be 0.02 and 0.04, respectively. The assay was found to be linear within the tested range of 0.063–111.6μg/L. The correlations between the second-generation Innotrac Aio!, Access, and AxSYM cTnI assays were good (rcoefficients 0.947–0.966), but involved differences in the measured concentrations, and the biases were highest with cTnI at low concentrations. The second-generation Innotrac Aio! cTnI assay was found to be superior to the first-generation assay with regard to precision in the low concentration range. The stability of the cTnI level was best in the serum, lithium-heparin plasma, and lithium-heparin whole blood samples (n=10, decrease<10%in 24 hours at+20°Cand at+4°C. There was no remarkable influence of hemodialysis on the cTnI release. False-positive cTnI values occurred in the presence of very high rheumatoid factor values, that is, over 3000 U/L. The 99th percentile of the apparently healthy reference group was0.03μg/L. The results demonstrate the very good analytical performance of the second-generation Innotrac Aio! cTnI assay.

Publisher

Hindawi Limited

Subject

Computer Science Applications,Instrumentation,Chemical Engineering (miscellaneous),Analytical Chemistry

Cited by 7 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3