Aims:
The aim of this retrospective study was to investigate SUVs variability with respect
to lesion size, administered dose, and reconstruction algorithm.
Background:
SUVmax and SUVpeak are influenced by technical factors as count statistics and reconstruction
algorithms.
Objective:
To fulfill the aim, we evaluated the SUVs variability with respect to lesion size, administered
dose, and reconstruction algorithm (ordered - subset expectation maximization plus point
spread function option - OSEM+PSF, regularized Bayesian Penalized Likelihood - BPL) in a 5 -
rings BGO PET/CT scanner.
Method:
Discovery IQ scanner (GE Healthcare, Milwaukee, Wisconsin, US) was used for list
mode acquisition of 25 FDG patients, 12 injected with 3.7 MBq/kg (Standard Dose protocol - SD)
and 13 injected with 1.8 MBq/kg (Low Dose protocol - LD). Each acquisition was reconstructed at
different time/FOV with both OSEM+PSF algorithm and BPL using seven different beta factors.
SUVs were calculated in 70 lesions and analysed in function of time/FOV and Beta. Image quality
was evaluated as a coefficient of variation of the liver (CV - liver).
Result:
SUVs were not considerably affected by time/FOV. However, SUVs were influenced by
beta: differences were higher in small lesions (37% for SUVmax, 15% for SUVpeak) compared to larger
ones (14% and 6%). CV - liver ranged from 6% with Beta-500 (LD and SD) to 13% with Beta-
200 (LD). CV - liver of BPL with Beta-350 (optimized for clinical practice in our institution) in
LD was lower than CV - liver of OSEM+PSF in SD.
Conclusion:
When a high sensitivity 5 - rings BGO PET/CT scanner is used with the same reconstruction
algorithm, quantification by means of SUVmax and SUVpeak is a robust standard compared
to the activity and scan duration. However, both SUVs and image quality are influenced by reconstruction
algorithms and the related parameters should be considered to obtain the best compromise
between detectability, quantification, and noise.