Affiliation:
1. Vocational School of Health Sciences, Department of Radiotherapy, Altınbaş University, Bakirkoy, Istanbul, Turkey
2. Medical Faculty, Department of Nuclear Medicine, Altınbaş University, Bakirkoy, Istanbul, Turkey
Abstract
Background:
New generation PET/CT devices provide quality images using low radiopharmaceutical
activities. Dose monitoring is carried out for nuclear medicine personnel, other
health personnel, and companions by determining the radiation dose emitted from low-activity
patients to the environment. In particular, it is necessary to revise the working conditions of the
personnel according to the radiation dose exposed.
Aims:
It was aimed to reevaluate the radiation dose rate transmitted to the environment from patients
injected with 18F-FDG.
Materials and Methods:
A total of 31 patients (14F, 17M) who underwent 18F-FDG PET/CT imaging
were included. The mean 18F-FDG activity of 7.26 ± 1.29 mCi was used for injection. After
injection, radiation dose rates (mR/h) were measured at distances of 25, 50, 100, 150, and 200cm
for 3 different periods from the level of the head, thorax, abdomen, and pelvis by using a GM
counter. Additionally, biological samples such as urine and sweat were taken during 3 different
periods. The activity amounts (μCi) in the samples were measured with a well-type counter.
Results:
Strong correlations were calculated between normalized dose rates obtained by all regions
and time. Considering the nuclear medicine staff handling time with a PET/CT patient, the
average dose received by staff was calculated between a range of 0.002-0.004 mSv/pt. The radiation
dose exposed to the porter and nurse was calculated as 0.049 mSv/pt for the 2nd hour and
0.001-0.007 mSv/pt for the 4th hour, respectively. The companion was exposed to a dose between
0.073-0.147 mSv and 0.024-0.048 mSv for public transport and private car transportation after 4-6
hours of injection (for 30-60 min of travel duration), respectively. For inpatients, the received dose
for porters, serving 20min from a distance of 30cm for the 2nd and 4th hours after the PET/CT scan,
was 0.049 mSv/pt and 0.048 mSv/pt, respectively. And for nurses serving from a 50cm distance
between 1-5 minutes, these values were found to be 0.001-0.007mSv/pt, 0.001-0.007mSv/pt, and
0.001-0.006mSv/pt, respectively.
Conclusion:
The radiation dose of nuclear medicine staff, porters, nurses, and companions are
found to be below the recommended dose limit by the ICRP. According to our results, there is no
need for any restrictions for patients, companions, or healthcare personnel in PET/CT units.
Funder
Altınbaş University Scientific Research Fund. Project
Publisher
Bentham Science Publishers Ltd.
Subject
Pharmacology,Radiology, Nuclear Medicine and imaging
Cited by
1 articles.