Author:
Doukky Rami,Rahaby Mouyyad,Alyousef Tareq,Vashistha Raj,Chawla Dave,Amin Amit P
Abstract
Purpose:
Soft-tissue attenuation patterns in SPECT-myocardial perfusion imaging (MPI) of supine acquisition systems are well recognized. Their prevalence and interaction with body-habitus and gender are ill-defined, which we sought to describe in this study.
Methods:
In a cross-sectional study, we described the prevalence of soft-tissue attenuation patterns in normal SPECT-MPI studies acquired with a supine patient-position SPECT system.
Results:
In 263 normal, clinically-indicated, supine-acquisition SPECT-MPIs the attenuation patterns observed were: anterior (35.4%), inferior (41.8%) and lateral (13.3%). Anterior attenuation was more prevalent among women (50.7% vs. 15.7%, P<0.001) and was associated with chest circumference among men. Conversely, inferior attenuation was more prevalent among men (78.3% vs. 13.5%, P<0.001) and was not affected by body-habitus. Lateral attenuation was more common among women (19.6% vs. 5.2%, p=0.001) and was associated with obesity (p=0.015).
Conclusions:
Soft-tissue attenuation artifacts are common in supine-acquisition SPECT-MPI. The recognition of their prevalence and association with body-habitus and gender is critical for the accurate interpretation of SPECT-MPI.