Author:
Chen Hsuan-Yu,Chen Chiachung
Abstract
Background:
Breath analysis can be used to screen disease and detect physical conditions. Many research studies have investigated various sensors or techniques that were used to detect gas concentrations and temperature. However, the studies were less concerned about the performance of sensors.
Methods:
In this study, we developed and tested a portable breath analyzer for exhaled gas measurement. We detected air flow rate, temperature, humidity, CO2 and O2 concentrations. Commercial sensing elements were used to assemble this breath analyzer. All sensors were calibrated to ensure their accuracy. The response times of different sensors were considered. We analyzed breath during three states; 1) seating and resting for 5 min, 2) walking for 5 min and 3) running for 5 min in test participants. Fifteen students 20 to 34 years of age were the test participants.
Results:
Heat, humidity and CO2 production and O2 consumption were calculated from these measured data and were used to assess the performance of the analyzer. The Respiratory Quotient (RQ) were estimated by the ratio of VCO2 to VO2. The RQ values for resting, walking and running states were 0.899, 0.865 and 0.785, respectively. Walking and running significantly increased the heat and humidity production of the human body, corresponding to the human physiological function.
Conclusion:
The RQ value is usually in the range of 0.8 to 0.85. The measurement results for our exhaled analyzer were close to this range. From the respiratory quotient values, the performance of this breath analyzer was reasonable.
Publisher
Bentham Science Publishers Ltd.
Subject
Biomedical Engineering,Medicine (miscellaneous),Bioengineering
Reference26 articles.
1. Lourenço C, Turner C.
Breath analysis in disease diagnosis: methodological considerations and applications.
Metabolites
2014;
4
(2)
: 465-98.
2. Prabhakar A, Iglesias RA, Shan X, et al.
Online sample conditioning for portable breath analyzers.
Anal Chem
2012;
84
(16)
: 7172-8.
3. Vincent TA, Wilson A, Hattersley JG, Chappell MJ, Gardner JW.
Development of a handheld side-streambreath analyser for point of care metabolicrate measurement.
2016.
In: Bioinformatics and Biomedical Engineering 4th International Conference, IWBBIO 2016;
Granada, Spain. 2016.
4. Popov TA, Dunev S, Kralimarkova TZ, Kraeva S, DuBuske LM.
Evaluation of a simple, potentially individual device for exhaled breath temperature measurement.
Respir Med
2007;
101
(10)
: 2044-50.
5. Popov TA, Kralimarkova T, Tzachev C, Dimitrov V, Mun KK, Gill J.
Exhaled breath temperature measurement made easy.
Pediatr Allergy Immunol
2009;
20
(2)
: 200-1.
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献