Affiliation:
1. Department of Applied Chemistry, University of Johannesburg, Doornfontein Campus, P.O. Box 17011, Johannesburg 2028, South Africa
Abstract
Background and Objectives:
Contamination of aquatic sediments by trace metals is one
of the global problems. This is because trace metals in sediments are persistent and nonbiodegradable.
They may pose danger to flora and fauna since they can be released into freshwater
systems. This study aimed at the development of microwave-assisted extraction using diluted hydrogen
peroxide and nitric acid for extraction of trace elements from sediment samples prior to inductively
coupled plasma optical emission spectrometry (ICP OES) determination.
Methods:
Response surface methodology (RSM) based on the Box-Behnken design was used for the
optimization of factors affecting the microwave-assisted extraction process. The optimum conditions,
for quantitative extraction of trace metals such as Cd2+, Cu2+, Cr2+, Pb2+ and Zn2+ were 16 min, 1.5 mol L-1
and 15% for extraction time, nitric acid concentration and H2O2 concentration, respectively.
Results and Discussion:
Under optimized conditions, the accuracy of the method was evaluated by
analyzing loamy clay certified reference materials (CRM052) and the recoveries were above 92%,
suggesting that the obtained results were in good agreement with the certified values. The developed
method has shown reproducibility (RSD < 5%), as well as relative low limits of detection (0.02-0.09 μg g-1)
and limit of quantitation (0.07-0.3 μg g-1). The developed analytical method was applied for extraction
and the determination of trace metals in freshwater sediment samples.
Conclusions:
The method displayed advantages such as simplicity, rapidity, environmentally friendly
and safe compared to classical methods that are based on concentrated acids.
Funder
Council for Geoscience and National Research Foundation
Publisher
Bentham Science Publishers Ltd.
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献