Torque Loss After Miniscrew Placement: An In-Vitro Study Followed by a Clinical Trial

Author:

Migliorati Marco,Drago Sara,Barberis Fabrizio,Schiavetti Irene,Dalessandri Domenico,Benedicenti Stefano,Biavati Armando Silvestrini

Abstract

To evaluate torque loss a week after insertion, both in an in vivo and an in vitro experimental setup were designed. In the in vivo setup a total of 29 miniscrews were placed in 20 patients who underwent orthodontic treatment. Maximum insertion torque (MIT) was evaluated at insertion time (T1). A week later, insertion torque was measured again by applying a quarter turn (T2); no load was applied on the screw during the first week. In the in vitro setup a total of 20 miniscrews were placed in pig rib bone samples. MIT was evaluated at insertion time (T1). Bone samples were kept in saline solution and controlled environment for a week during which the solution was refreshed every day. Afterwards, torque was measured again by applying a quarter turn (T2). The comparison of MIT over time was done calculating the percentage difference of the torque values between pre- and post-treatment and using the parametric two independent samples t-test or the non-parametric Mann–Whitney test. After a week unloaded miniscrews showed a mean loss of rotational torque of 36.3% and 40.9% in in vitro and in in vivo conditions, respectively. No statistical differences were found between the two different setups. Torque loss was observed after the first week in both study models; in vitro experimental setup provided a reliable study model for studying torque variation during the first week after insertion.

Publisher

Bentham Science Publishers Ltd.

Subject

General Dentistry

Reference30 articles.

1. Kanomi R. Mini-implant for orthodontic anchorage. J Clin Orthod 1997; 31 (11) : 763-7.

2. Costa A, Raffainl M, Melsen B. Miniscrews as orthodontic anchorage: a preliminary report. Int J Adult Orthodon Orthognath Surg 1998; 13 (3) : 201-9.

3. Kyung HM, Park HS, Bae SM, Sung JH, Kim IB. Development of orthodontic micro-implants for intraoral anchorage. J Clin Orthod 2003; 37 (6) : 321-8.

4. Sung JH, Kyung HM, Bae SM, Park HS, Kwon OW, McNamara J. Microimplants in Orthodontics. Daegu: Dentos Inc 2006.

5. Mesa F, Muñoz R, Noguerol B, de Dios Luna J, Galindo P, O’Valle F. Multivariate study of factors influencing primary dental implant stability. Clin Oral Implants Res 2008; 19 (2) : 196-200.