Abstract
Background:
Due to their increased precision, CAD/CAM generated bars (Computer-Aided Design/ Computer-Aided Manufacturing) are increasingly utilized in implant prosthodontics. For optimal clinical results, surface morphology should promote the integration of soft tissue while minimizing plaque and bacterial retention.
Objective:
Despite their clinical use, only limited information on the biological and clinical surface quality of CAD/CAM milled bars is available. The aim of the study was therefore to characterize the surface topography of bars of different manufacturers based on the profilometric analysis and the need for manual post-processing in the laboratory.
Methods:
A custom mandibular edentulous cast with four anterior implants was used as a reference cast and reproduced eight times. On each reproduction cast, corresponding scan flags were positioned and digitized. Acrylic 3D printed bar frameworks were produced and sent to the respective production center along with the digital files of the CAD bars for milling. In the course of profilometric analysis, all bars were examined in three critical Regions of Interest (ROI): Transmucosal, labial, basal. Sa and Ra values of each construction were determined. To evaluate the necessary refinishing time eight dental technicians macroscopically evaluated the bars by performing a subjective visual inspection. Kruskal-Wallis H-tests and Tukey and Kramer's post hoc tests were applied to detect differences between the samples.
Results:
After profilometric examination, three specimens (Dentsply Sirona: ZDC; Straumann: ZST; CAMLOG: ZCC) demonstrated surface roughness values in the biological acceptable range (Sa 0.2-0.4 μm) in the transmucosal region and provided optimal conditions for a reliable soft tissue adaptation. The Ra measurements revealed values beyond the acceptable threshold in the transmucosal region for three bars (Straumann: ZST; Dentsply Sirona: ZDC; Amann Girrbach: LAC). Four bars (LAC: Amann Girrbach; ZBC: BEGO; Datron: LDC & LDT; Zirkonzahn: ZZC) needed undesirable extensive manual rework. The evaluation of quality and time for manual post-processing by dental technicians confirmed the measurement-based ranking of the bars.
Conclusion:
It is desirable to define a clear roughness threshold for the clinical acceptance of transmucosal CAD/CAM generated surfaces. Clinical studies with profilometric data could help to further improve the surface quality of CAD/CAM milled bars and reduce the need for manual reworking time and effort.
Publisher
Bentham Science Publishers Ltd.
Reference54 articles.
1. Rinke S, Rasing H, Gersdorff N, Buergers R, Roediger M.
Implant-supported overdentures with different bar designs: A retrospective evaluation after 5-19 years of clinical function.
J Adv Prosthodont
2015;
7
(4)
: 338-43.
2. Rentsch-Kollar A, Huber S, Mericske-Stern R.
Mandibular implant overdentures followed for over 10 years: Patient compliance and prosthetic maintenance.
Int J Prosthodont
2010;
23
(2)
: 91-8.
3. Meijer HJ, Raghoebar GM, Batenburg RH, Vissink A.
Mandibular overdentures supported by two Brånemark, IMZ or ITI implants: A ten-year prospective randomized study.
J Clin Periodontol
2009;
36
(9)
: 799-806.
4. Meijer HJ, Raghoebar GM, Batenburg RH, Visser A, Vissink A.
Mandibular overdentures supported by two or four endosseous implants: A 10-year clinical trial.
Clin Oral Implants Res
2009;
20
(7)
: 722-8.
5. Dantas IdeS, Souza MB, Morais MH, Carreiro AdaF, Barbosa GA.
Success and survival rates of mandibular overdentures supported by two or four implants: A systematic review.
Braz Oral Res
2014;
28
: 74-80.
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献