The proper course and reproducibility of diagnostic techniques depend on narrowly defined reaction conditions,
including the reaction pH. Nevertheless, numerous assays are affected by an inaccurately defined reaction pH.
Buffers are sometimes suggested for use outside their useful pH ranges, which complicates the reproducibility of results
because the buffering capacity is insufficient to retain the disclosed pH. Here, we focus on the comet assay lysis
buffer. Comet assay is broadly used for quantifying DNA breaks in eukaryotic cells. The most widespread comet assay
protocols employ lysis of the cells before electrophoresis in a buffer containing Triton X-100, a high concentration of
NaCl, sodium sarcosinate, EDTA, and Tris, with some modifications. However, nearly all researchers report that they
use Tris buffer at pH 10, and some report the pH of the Tris additive alone. Alternatively, others report the pH of the
final lysis buffer. However, the lysis solution used in the comet assay is buffered at a pH outside the useful range of
Tris. Tris-based buffers have a useful pH range of 7.0 - 9.0. The buffer composed of 10 mM Tris has pKa 8.10 at 25°C
and 8.69 at 4°C. The cell lysis conditions used in nearly all modifications of comet assay protocols remain imprecise
and uncritically employed. Despite the pH of the lysis buffer likely has negligible effect on the detection of DNA
breaks, precise lysis conditions are highly important for the use of comet assay in the detection of base modifications,
which are often unstable and sensitive to pH.