Affiliation:
1. Department of Cardiology, Harran University, Sanliurfa, Turkey
2. Department of Medical Biochemistry, Harran University, Sanliurfa, Turkey
Abstract
Background:
The discovery that biotin interferes with results of troponin and Nt-proBNP led
some commercial firms to update their measurement methods. In particular, the clinical incompatibility of
cardiac test results may affect the risk of morbidity and mortality.
Objective:
The aim of this study is to investigate the interference effects of 7 different contrast agents on
cardiac markers (Troponin-I, Nt-proBNP, Mass CK-MB, CK, AST, LDH) and in coagulation tests (PT,
APTT).
Methods:
Seven different contrast media were added into control materials by using interference protocol.
The concentrations of PT, APTT, CK, AST, LDH, Mass CK-MB, Troponin-I, and Nt-proBNP were
measured by Sysmex CS-2100, Abbott c16000, Siemens Centaur XP and AFİAS-6 analyzer. The number
of deviations from target values was calculated.
Results:
The 7 different contrast media caused negative interference in troponin levels between 57.43%
and 62.87%. It was found that different contrast media produced false negativity in the Nt-proBNP test,
ranging from 6.11% to 96.01%. Enzymes and coagulation tests were less affected.
Conclusion:
Different contrast media may cause false negatives in cTnI and Nt-proBNP. The contrast
medium that causes the least interference should be preferred. The results of samples taken in the first hour
after contrast imaging should be interpreted with care.