Affiliation:
1. Department of Mechanical Engineering, University of Leeds
2. DePuy International, Leeds
Abstract
A number of studies of explanted metallic femoral heads have shown scratches or damage caused by bone cement, bone or metallic particles. This damage has been cited as a cause of increased wear of ultra-high molecular weight polyethylene (UHMWPE) acetabular cups. In this laboratory study, small scratches 2 μm deep were made on smooth stainless steel surfaces at a spacing of 10 mm. These individual scratches were found to increase the wear rate of UHMWPE by a factor of 30 in unidirectional sliding and a factor of 70 in reciprocating motion. It is of particular concern that a single small scratch, which is not detected by the average surface roughness measurement Ra can cause such a dramatic increase in the wear of ultra-high molecular weight polyethylene.
Subject
Mechanical Engineering,General Medicine
Cited by
122 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献