Affiliation:
1. Diabetic Department, King's College London and King's College Hospital
2. Department of Medical Engineering and Physics, King's College London
Abstract
Thin pressure-measuring insoles have enabled new clinical studies of in-shoe distribution of plantar pressure. One of these new devices, based on force-sensitive resistor (FSR) technology, exhibits time-varying behaviour due to the mechanical properties of the sensor materials. This has been variously attributed to initial bedding in, wear or time-dependency behaviour. Cyclic and sustained loading were used in laboratory conditions to investigate this phenomenon. The results show that the sensitivity of the sensor varies rapidly over the first minute of loading. There is also a slower long-term effect which reduces over time. From this information, procedures for optimal calibration and use are suggested.
Subject
Mechanical Engineering,General Medicine
Cited by
17 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献