Mechanistic and Morphological Origins of Ultra-High Molecular Weight Polyethylene Wear Debris in Total Joint Replacement Prostheses

Author:

Wang A1,Stark C1,Dumbleton J H1

Affiliation:

1. Howmedica Incorporated, Division of Pfizer Hospital Products Group, Rutherford, New Jersey, USA

Abstract

The mechanistic and morphological origins of microscopic wear debris generated from UHMWPE articular surfaces in total joint replacement prostheses are investigated in this study. It was found experimentally that the molecular chain structure at the articulating surface of UHMWPE undergoes a re-organization process due to strain accumulation caused by surface traction. This molecular re-organization process creates a fibre-like surface texture that exhibits an anisotropic behaviour similar to a unidirectionally reinforced polymer composite. This composite responds to stresses favourably if loaded along the fibre axis but unfavourably if loaded off axis. Due to the very complex multi-axial motion/loading nature at the articular surfaces in total joint replacements, the stress tensors applied to each localized asperity contact area continuously change their directions and magnitudes. These changes in the localized stress field create an off-axis loading situation at each localized contact zone with respect to the orientation of the molecular chains. Depending on the off-axis angle, failure of the molecular structure occurs in three different ways: tensile rupture at very small off-axis angles, shear rupture at intermediate off-axis angles and transverse splitting at large off-axis angles. These failure mechanisms all produce similar fibre-like wear debris. However, the failure stresses differ significantly among the three modes. According to this molecular wear theory, the preferred polymer microstructure for optimal wear resistance would be a three-dimensionally strong network connected by covalent bonds between molecular chains. For UHMWPE, a three-dimensional molecular network can be created by radiation induced cross-linking. Experiments conducted on both gamma irradiated and unirradiated UHMWPE specimens using a linear wear machine and multi-axial joint simulators confirmed the validity of the molecular wear theory.

Publisher

SAGE Publications

Subject

Mechanical Engineering,General Medicine

Cited by 236 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3