Studies of explanted hip prostheses have shown high wear rates of ultra-high molecular weight polyethylene (UHMWPE) acetabular cups and roughening of the surface of the metallic femoral head. Bone and bone cement particles have also been found in the articulating surfaces of some joints. It has been proposed that bone or bone cement particles may cause scratching and deterioration in the surface finish of metallic femoral heads, thus producing increased wear rates and excessive amounts of wear debris. Sliding wear tests of UHMWPE pins on stainless steel have been performed with particles of different types of bone and bone cement added. Damage to the stainless steel counterface and the motion of particles through the interface have been studied. Particles of bone cement with zirconium and barium sulphate additives and particles of cortical bone scratched the stainless steel counterface. The cement particles with zirconium additive produced significantly greater surface damage. The number of particles entering the contact and embedding in the UHMWPE pin was dependent on particle size and geometry, surface roughness and contact stress. Particles are likely to cause surface roughening and increased wear rates in artificial joints.