Affiliation:
1. University College London and Royal National Orthopaedic Hospital Trust, Stanmore, Middlesex Department of Biomedical Engineering UK
Abstract
The starting point of this article is a general design criterion applicable to all types of total knee replacement. This criterion is then expanded upon to provide more specifics of the required kinematics, and the forces which the total knee must sustain. A characteristic which differentiates total knees is the amount of constraint which is required, and whether the constraint is translational or rotational. The different forms of total knee replacement are described in terms of these constraints, starting with the least constrained unicompartments to the almost fully constrained fixed and rotating hinges. Much attention is given to the range of designs in between these two extreme types, because they constitute by far the largest in usage. This category includes condylar replacements where the cruciate ligaments are preserved or resected, posterior cruciate substituting designs and mobile bearing knees. A new term, ‘guided motion knees’, is applied to the growing number of designs which control the kinematics by the use of intercondylar cams or specially shaped and even additional bearing surfaces. The final section deals with the selection of an appropriate design of total knee for specific indications based on the design characteristics.
Subject
Mechanical Engineering,General Medicine
Cited by
48 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献