Comparative repeatability of guide-pin axis positioning in computer-assisted and manual femoral head resurfacing arthroplasty

Author:

Hodgson A12,Helmy N3,Masri B A3,Greidanus N V3,Inkpen K B1,Duncan C P3,Garbuz D S3,Anglin C1

Affiliation:

1. Department of Mechanical Engineering, University of British Columbia, Vancouver, Canada

2. Institute for Computing, Information and Cognitive Systems, University of British Columbia, Vancouver, Canada

3. Department of Orthopaedics, University of British Columbia, Vancouver, Canada

Abstract

The orientation of the femoral component in hip resurfacing arthroplasty affects the likelihood of loosening and fracture. Computer-assisted surgery has been shown to improve significantly the surgeon's ability to achieve a desired position and orientation; nevertheless, both bias and variability in positioning remain and can potentially be improved. The authors recently developed a computer-assisted surgical (CAS) technique to guide the placement of the pin used in femoral head resurfacing arthroplasty and showed that it produced significantly less variation than a typical manual technique in varus/valgus placement relative to a pre-operatively determined surgical plan while taking a comparable amount of time. In the present study, the repeatability of both the CAS and manual techniques is evaluated in order to estimate the relative contributions to overall variability of surgical technique (CAS versus manual), surgeon experience (novice versus experienced), and other sources of variability (e.g. across specimens and across surgeons). This will enable further improvements in the accuracy of CAS techniques. Three residents/fellows new to femoral head resurfacing and three experienced hip arthroplasty surgeons performed 20-30 repetitions of each of the CAS and manual techniques on at least one of four cadaveric femur specimens. The CAS system had markedly better repeatability (1.2°) in varus/valgus placement relative to the manual technique (2.8°), slightly worse repeatability in version (4.4° versus 3.2°), markedly better repeatability in mid-neck placement (0.7 mm versus 2.5 mm), no significant dependence on surgeon skill level (in contrast to the manual technique), and took significantly less time (50 s versus 123 s). Proposed improvements to the version measurement process showed potential for reducing the standard deviation by almost two thirds. This study supports the use of CAS for femoral head resurfacing as it is quicker than the manual technique, independent of surgeon experience, and demonstrates improved repeatability.

Publisher

SAGE Publications

Subject

Mechanical Engineering,General Medicine

Cited by 27 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3