Influence of cup orientation on the wear performance of metal-on-metal hip replacements

Author:

Angadji A1,Royle M1,Collins S N2,Shelton J C1

Affiliation:

1. School of Engineering and Materials Science, Queen Mary University of London, London, UK

2. Corin Ltd, Cirencester, UK

Abstract

There are a number of factors that determine the overall outcome of total hip replacement (THR) surgery, some of which appear to be related to the surgical procedure. In particular, the inclination angle at which the acetabular component is placed has been reported to influence the long-term successful performance of THR. The present study assessed the influence of cup orientation on the wear of 40 mm diameter metal-on-metal (MoM) hip bearings tested in a hip simulator. The bearings had a mean radial clearance of 150 μm; the cups oriented at 35°, 50°, and 60° to the horizontal were loaded for up to 6×106 cycles. In each test the wear rates during the run-in phase were higher than in the steady state phase; the wear rates during the run-in phase were not significantly different for each cup orientation. However, at cup angles of 50° and 60°, the steady state wear rates were 0.69 mm3/106 cycles and 1.7 mm3/106 cycles respectively, significantly higher than at 35° (0.24 mm3/106 cycles). The results indicated that larger cup inclination angles not only move the position of the wear scar but also, more significantly in MoM bearings, increase the wear rates and total wear volume generated.

Publisher

SAGE Publications

Subject

Mechanical Engineering,General Medicine

Cited by 82 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3