Abstract
A basic problem common to all systems for computer assisted surgery (CAS) is patient referencing, or the transfer of preoperative image data to the intraoperative pathology. The authors describe a highly precise CAS system with non-invasive referencing that can be used in ear, nose and throat (ENT) surgery of the paranasal sinuses. It is based on optical digitizing with several custom-made self-localizing surgical instruments. The accuracy of the system was tested in an experimental model using a plastic head. Measurements of repositioning the reference bow had a mean error of 0.81 mm + 0.31 mm. The system was evaluated clinically with 11 patients who received surgery for different pathologies of the paranasal sinuses. These trials met with a high rate of success and specific results are reported.