Measurement of pulsatile haemodynamic forces in a model of a bifurcated stent graft for abdominal aortic aneurysm repair

Author:

Zhou S N1,How T V1,Black R A1,Vallabhaneni S R2,McWilliams R3,Brennan J A2

Affiliation:

1. Division of Clinical Engineering, School of Clinical Sciences, University of Liverpool, Liverpool, UK

2. Regional Vascular Surgery Unit, Royal Liverpool University Hospital Trust, Liverpool, UK

3. Department of Radiology, Royal Liverpool University Hospital Trust, Liverpool, UK

Abstract

The longitudinal haemodynamic force (LF) acting on a bifurcated stent graft for abdominal aortic aneurysm repair has been estimated previously using a simple one-dimensional analytical model based on the momentum equation which assumes steady flow of an inviscid fluid. Using an instrumented stent-graft model an experimental technique was developed to measure the LF under pulsatile flow conditions. The physical stent-graft model, with main trunk diameter of 30 mm and limb diameters of 12 mm, was fabricated from aluminium. Strain gauges were bonded on to the main trunk to determine the longitudinal strain which is related to the LF. After calibration, the model was placed in a pulsatile flow system with 40 per cent aqueous glycerol solution as the circulating fluid. The LF was determined using a Wheatstone bridge signal-conditioning circuit. The signals were averaged over 590 cardiac cycles and saved to a personal computer for subsequent processing. The LF was strongly dependent on the pressure but less so on the flowrate. The measured forces were higher than those predicted by the simplified mathematical model by about 6–18 per cent during the cardiac cycle. The excess measured forces are due to the viscous drag and the effect of pulsatile flow. The peak measured LF in this model of 30 mm diameter may exceed the fixation force of some current clinical endovascular stent grafts.

Publisher

SAGE Publications

Subject

Mechanical Engineering,General Medicine

Cited by 8 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3