Influence of contaminants in the stem—ball interface on the static fracture load of ceramic hip joint ball heads

Author:

Weisse B1,Affolter C1,Stutz A1,Terrasi G P1,Köbel S2,Weber W2

Affiliation:

1. Laboratory for Mechanical Systems Engineering, Empa — Swiss Federal Laboratories for Materials Testing and Research, Dübendorf, Switzerland

2. Metoxit AG, Thayngen, Switzerland

Abstract

The probability of in-vivo failure of ceramic hip joint implants is very low (0.05–0.004 per cent). Besides material flaws and overloading, improper handling during implantation may induce fractures of the ceramic ball head in the long term. This study focuses on the influence of contaminants located in the stem—ball interface and on the use of damaged metal tapers on the strength of ceramic ball heads. Mechanical tests on alumina ball heads according to the standard ISO 7206-10 were performed to identify their effect on the static fracture load. A decrease of up to 90 per cent with respect to the reference static fracture load was found when contaminants such as bone chips, soft tissue, or blood were present. Reductions of 57 per cent and 27 per cent were observed for deformed stem cross-sections (from circular to elliptical) and for flattened stems respectively, making deformed stems another influential parameter. Since any alteration of the interface between the metal taper and the ceramic ball head yields a non-uniform load introduction and hence results in stress concentrations, its presence has to be avoided.

Publisher

SAGE Publications

Subject

Mechanical Engineering,General Medicine

Reference21 articles.

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3