Author:
Koerber Raphaelle M.,Vaccarello Linh,Ho Allan
Abstract
BACKGROUND
This study evaluated the effectiveness of the reverse stethoscope technique in improving speech intelligibility. In this technique, a clinician places the earpieces of their stethoscope into the ears of a hearing-impaired patient and speaks into the chest piece.
METHODS
The International Speech Test Signal was presented to four Littman® stethoscope models and a Pocketalker® personal voice amplifier using an Audioscan® hearing instrument test box. The acoustic outputs of the stethoscopes and voice amplifier were measured across the frequency spectrum of speech. The Speech Intelligibility Index of the resulting speech was calculated for natural speech and for each device in relation to 10 standardized hearing losses representing the population of older adults.
RESULTS
For each of the 10 hearing losses, the speech signal emitted by the stethoscopes was quieter and yielded lower speech intelligibility scores than regular speech. In contrast, the voice amplifier provided mid and high frequency amplification and improved speech intelligibility scores for all but the mildest hearing losses.
CONCLUSIONS
The reverse stethoscope technique worsens the clarity of speech and should not be used with older, hearing-impaired patients. Instead, clinicians should use regular speech or, preferably, personal voice amplifiers.
Publisher
Canadian Geriatrics Society
Subject
Geriatrics and Gerontology,Gerontology
Cited by
3 articles.
订阅此论文施引文献