Abstract
Objective: To compare the trending ability, accuracy, and precision of non-invasive stroke volume (SV) measurement based on a bioreactance technique and measurement of the pulse wave transit time (PWTT) versus the esophageal Doppler monitoring (EDM).
Materials and Methods: Two hundred twenty-seven paired measurements from 10 patients who underwent abdominal surgery under general anesthesia were included for SV measurements. Pearson’s correlation coefficient was calculated, and Bland-Altman analysis was performed to evaluate the agreement between EDM and bioreactance (EDM-bioreactance) and between EDM and PWTT (EDM-PWTT).
Results: EDM-bioreactance had a correlation coefficient of 0.75 (95% confidence interval [CI] 0.62 to 0.78; p<0.001), bias of 0.28 ml (limits of agreement –30.92 to 31.38 ml), and percentage error of 46.82%. EDM-PWTT had a correlation coefficient of 0.48 (95% CI 0.44 to 0.72; p<0.001), bias of –0.18 ml (limits of agreement –40.28 to 39.92 ml), and percentage error of 60.17%. A subgroup analysis of data from patients who underwent crystalloid loading was performed to detect the trending ability. The four-quadrant plot analysis between EDM-bioreactance and EDM-PWTT demonstrated concordance rates of 70.00% and 73.68%, respectively.
Conclusion: SV measurement based on bioreactance technique and measurement of PWTT are not interchangeable with EDM.
Trial registration: Thai Clinical Trials Registry, TCTR 20181217003
Keywords: Stroke volume, Cardiac output, Doppler, Perioperative care, Pulse, Time
Publisher
Medical Association of Thailand
Reference23 articles.
1. Aya HD, Cecconi M, Hamilton M, Rhodes A. Goal-directed therapy in cardiac surgery: a systematic review and meta-analysis. Br J Anaesth 2013;110:510-7.
2. Giglio M, Manca F, Dalfino L, Brienza N. Perioperative hemodynamic goal-directed therapy and mortality: a systematic review and meta-analysis with meta-regression. Minerva Anestesiol 2016;82:1199-213.
3. Pearse RM, Harrison DA, MacDonald N, Gillies MA, Blunt M, Ackland G, et al. Effect of a perioperative, cardiac output-guided hemodynamic therapy algorithm on outcomes following major gastrointestinal surgery: a randomized clinical trial and systematic review. JAMA 2014;311:2181-90.
4. Sun Y, Chai F, Pan C, Romeiser JL, Gan TJ. Effect of perioperative goal-directed hemodynamic therapy on postoperative recovery following major abdominal surgery-a systematic review and meta-analysis of randomized controlled trials. Crit Care 2017;21:141.
5. Binanay C, Califf RM, Hasselblad V, O’Connor CM, Shah MR, Sopko G, et al. Evaluation study of congestive heart failure and pulmonary artery catheterization effectiveness: the ESCAPE trial. JAMA 2005;294:1625-33.