Multicenter clinical validation of an on-line monitor of dialysis adequacy.

Author:

Depner T A,Keshaviah P R,Ebben J P,Emerson P F,Collins A J,Jindal K K,Nissenson A R,Lazarus J M,Pu K

Abstract

Quantitation of hemodialysis by measuring changes in blood solute concentration requires careful timing when taking the postdialysis blood sample to avoid errors from postdialysis rebound and from recirculation of blood through the access device. It also requires complex mathematical interpretation to account for solute disequilibrium in the patient. To circumvent these problems, hemodialysis can be quantified and its adequacy assessed by direct measurement of the urea removed in the dialysate. Because total dialysate collection is impractical, an automated method was developed for measuring dialysate urea-nitrogen concentrations at frequent intervals during treatment. A multicenter clinical trial of the dialysate monitoring device, the Biostat 1000 (Baxter Healthcare Corporation, McGaw Park, IL) was conducted to validate the measurements of urea removed, the delivered dialysis dose (Kt/V), and net protein catabolism (PCR). The results were compared with a total dialysate collection in each patient. During 29 dialyses in 29 patients from three centers, the paired analysis of urea removed, as estimated by the dialysate monitor compared with the total dialysate collection, showed no significant difference (14.7 +/- 4.7 g versus 14.8 +/- 5.1 g). Similarly, measurements of Kt/V and PCR showed no significant difference (1.30 +/- 0.18 versus 1.28 +/- 0.19, respectively, for Kt/V and 42.3 +/- 15.7 g/day versus 52.2 +/- 17.4 g/day for PCR). When blood-side measurements during the same dialyses were analyzed with a single-compartment, variable-volume model of urea kinetics, Kt/V was consistently overestimated (1.49 +/- 0.29/dialysis, P < 0.001), most likely because of failure to consider urea disequilibrium. Because urea disequilibrium is difficult to quantitate during each treatment, dialysate measurements have obvious advantages. The dialysate monitor eliminated errors from dialysate bacterial contamination, simplified dialysate measurements, and proved to be a reliable method for quantifying and assuring dialysis adequacy.

Publisher

American Society of Nephrology (ASN)

Subject

Nephrology,General Medicine

Cited by 58 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3