Abstract
Introduction: Currently, several molecular assays are available to detect and quantify HBV DNA in clinical samples. We aimed to characterize and compare the clinical performance of newly designed NeuMoDx PCR to the existing artus PCR.
Methodology: The plasma HBV DNA levels of 96 clinical and 5 external quality control samples were measured by NeuMoDx and artus assays simultaneously in Kocaeli University, Turkey. The linearity, agreement and the correlation between two assays were determined by Deming regression analysis, Bland-Altman plotting, the chi-square and the relative absolute error statistical analyzes. For all statistical analyzes, the XLSTAT statistical program was used.
Results: The mean (standard deviation; SD) age was 45.07 ± 12.29. HBsAg S/Co median (range) was 4,273.4 ± 1,138.1 and ALT U/L median (range) was 27 ± 16. The mean (SD) of HBV DNA was 1.46+E6 ± 1.0+E4 for NeuMoDx and 1.54+E5 ± 4.7 + E4 for artus assays. The Deming regression indicates a linear correlation (95% confidence). The chi-square test indicates strong correlation (p < 0.001). Bland-Altman analysis confirms that the measurement difference is acceptable. The relative absolute error analysis for artus showed relatively less and more consistent error rate. With 5 external quality check samples, the statistical significance was low (p = 0.566).
Conclusions: The NeuMoDx HBV assay showed an excellent analytical performance by providing a rapid, high throughput technology in a random-access testing system in clinical samples and may be a new solution for viral load quantification in the management of HBV infections.
Publisher
Journal of Infection in Developing Countries
Subject
Virology,Infectious Diseases,General Medicine,Microbiology,Parasitology
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献