Abstract
The Woven EndoBridge (WEB) (MicroVention/Terumo) device is a treatment option for wideneck bifurcation aneurysms. An uncommon adverse effect is WEB device migration. While certain bailout strategies for WEB recovery have been described, there is still a paucity of information on optimal strategies to maximize both short and long-term post-operative outcomes. We add 2 cases at our institution to the existing literature of WEBectomy in the setting of complicated intracranial aneurysm treatment. We discuss the long-term imaging outcomes with additional fluoroscopy video demonstrating our technique. Our findings reflect a clear benefit for the use of the Amplatz Gooseneck<sup>TM</sup> microsnare (Medtronic) device as a means of WEB recovery, coupled with potential stent-assisted WEB embolization to remove the aneurysm from the parent circulation, while minimizing recurrence and thromboembolic complications.
Publisher
Korean Society of Interventional Neuroradiology
Subject
Radiology, Nuclear Medicine and imaging,Neurology (clinical)
Cited by
1 articles.
订阅此论文施引文献