Abstract
A flow diverter can be used to treat a ruptured aneurysm, especially during complicated cases for endovascular or open microsurgical procedures. However, some complications have been reported such as occlusion of side branches or perforating arteries, stent migration or improper stent position, re-rupture of the aneurysm, and in-stent thrombosis. A 38-year-old man diagnosed with a ruptured left supraclinoid aneurysm was treated with a Pipeline embolization device. Despite adequate stent coverage of the aneurysmal neck, delayed proximal migration was seen in follow-up imaging. To our knowledge, proximal migration of the stent mostly happens intraoperatively due to an undersized stent or excessive stretching. We report a case of delayed proximal flow diverter stent migration.