Ionometric sensor for express determination of hexamethylenetetramine

Author:

Volnyanska O.V., ,Mironyak M.O.,Labyak O.V.,Nikolenko M.V., ,

Abstract

The article is devoted to the development of an ionometric sensor for the determination of hexamethylenetetramine (HMTA). It was found that poorly soluble complex compounds with the composition (HMTA)3(PMo12O40)2 and (HMTA)3(PW12O40)2 are formed as a result of the interaction of the cation of the test substance with the anions of heteropoly acids. The obtained compounds were used as electrode active substances for polyvinyl chloride membranes of ionometric sensors. Phthalic acid derivatives (dibutyl and dioctyl phthalates) were used as solvents for plasticized membranes. The optimal composition of the membranes was established. The dependences of the electrochemical properties of the obtained sensor membranes on various factors were studied. This made it possible to choose the optimal conditions for the operation of the ionometric sensor. The response time of the sensor does not exceed 3 minutes, and the service life of the membrane (about 40 days) allows performing determination without replacement. The detection limit of hexamethylenetetramine using this sensor is 1.010–5 mol/l, and quantitative determination can be carried out in a fairly short time (5–10 min).

Publisher

SHEI Ukrainian State University of Chemical Technology

Subject

Materials Chemistry,General Chemical Engineering,Environmental Chemistry,General Chemistry

Reference10 articles.

1. 1. Sukhorukov AYu, Ioffe SL. Urotropin: dremavshaya istoriya i probuzhdenie [Urotropine: dormant history and awakening]. Vestnik RFFI. 2017; 4(96): 20-27. (in Russian). doi: 10.22204/2410-4639-2017-096-04-20-31.

2. 2. Nechaev AP, Kochetkova AA, Zaitsev AN. Pishchevye dobavki [Food additives]. Moscow: Kolos-Press; 2002. 256 p. (in Russian).

3. 3. Karakhanyan ZKh, Korzanov VS. Opredelenie soderzhaniya geksametilentetramina v vodnykh rastvorakh metodom konduktometricheskogo titrovaniya [The determination of hexamethylenetetramine content in aqueous solutions by conductometric titration method]. Vestn Permskogo Univer Ser Khim. 2017; 7(2): 159-166. (in Russian). doi: 10.17072/2223-1838-2017-2-159-166.

4. 4. Ponomarenko SP, Borovikov YuYa, Sivachek TE, Makovetskiy VP. Elektricheskie svoistva urotropina i ego kompleksov s protonodonarami [Electrical properties of urotropine and its complexes with donors of protons]. Ukr Khim Zhurn. 2008; 74(2): 92-96. (in Russian).

5. 5. Tininis AG, Leandro A, Pezza HR, Melios CB, Pezza L. Rapid spot test analysis for the detection of urotropine in pharmaceutical preparations. Anal Lett. 2000; 33(14): 2901-2912. doi: 10.1080/00032710008543229.