The number of procedures for upgrading implantable devices for cardiac resynchronization therapy has increased considerably during the last decade. A major challenge that operators face in these circumstances is occlusion of the access vein. We have modified a pull-through method to overcome this obstacle.
Six consecutive patients with occluded access veins and well-developed collateral networks underwent a procedure in which the occluded vein was recanalized by snaring the existing atrial lead via transfemoral access. Upgrading the device was successful in all patients; none had intraprocedural complications. Our experience shows that our modified pull-through technique may be a feasible alternative for upgrading cardiac resynchronization therapy in patients with venous occlusion.