Author:
Green Kelly D.,Waked Alain,Majeed Usman,Fudim Marat,Robbins Mark A.,Crenshaw Marshall,Zhao David
Abstract
We report the case of an 85-year-old woman with severe aortic stenosis who underwent transcatheter aortic valve replacement with use of the Edwards Sapien® valve system. The procedure was complicated by rupture of the valve-deployment balloon, with separation and retention of the nose cone of the RetroFlex 3® delivery system in the iliac artery. Our endovascular retrieval of the equipment was successful, and we achieved access-site hemostasis by deploying a covered stent. To our knowledge, this is the first report of the endovascular retrieval of a malfunctioning delivery system during transcatheter aortic valve replacement.