Extracorporeal ultrasound exposure by the low-frequency acoustic amplitude-modulated signal on a ureteral stent for preventing its incrustation: experimental determination of optimal application points

Author:

Tsukanov A. Yu.1ORCID,Akhmetov D. S.1ORCID,Novikov A. A.2ORCID,Negrov N. A.2ORCID,Putintseva A. R.2ORCID,Semikina S. P.1ORCID

Affiliation:

1. Omsk State Medical University

2. Omsk State Technical University

Abstract

Introduction. Incrustation and biofilms formation on the surface of ureteral stents are still the most significant complications of internal drainage of the upper urinary tract. There are much researchers conducted to combat these complications. The lack of a solution to this problem affects the ultimate results of treatment and economic losses. The issue of impact by physical methods on the ureteral stent, particularly the use of extracorporeal ultrasound acoustic exposure remains, promising and poorly covered.Purpose of the study. To determine the optimal application points of extracorporeal acoustic exposure by the low-frequency ultrasonic amplitude-modulated signal on a ureteral stent in an experiment.Materials and methods. The original device was designed. The main principle of its operation is the generation of an amplitude-modulated ultrasonic signal in two modes: pulsed and permanent. A sexually mature mongrel dog was an experimental animal. The ureteral stent was placed by laparotomy and cystotomy. Intraoperatively, the emitter of the developed device was applied to the skin of the animal, according to the previously indicated topographic and anatomical landmarks. At the same time, an ultrasonic wave noise analyzer was applied through the laparotomy wound to the appropriate level of the ureter. Measurements of ultrasound intensity indicators were performed three times in two operating modes of the device.Results. Pulsed mode: for the ureteral upper third, the highest ultrasonic intensity (123.67 dB) was achieved along the posterior axillary line. For the ureteral middle third, the best ultrasound intensity (115 dB) was obtained by the posterior axillary line. For the ureteral lower third, the highest ultrasound intensity (113.67 dB) was noted along the middle axillary line.Permanent mode: the best ultrasonic intensity in the projection of the ureteral upper, middle, and lower thirds was achieved along the posterior axillary line and was 118.67 dB, 117 dB and 116.67 dB, accordingly. However, there was an excessive heat effect, manifested by hyperemia and hyperthermia of the animal's skin, fascicular muscle contractions during the instrument functioned in the permanent mode, which can potentially lead to thermal burns and intolerance to the procedure.Conclusion. The pulsed mode of the device function is most safe. The optimal application points of the instrument emitter for the ureteral upper and middle thirds is the posterior axillary line, and for the ureteral lower third is the middle axillary line.

Publisher

Rostov State Medical University

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3