Determining the error in measuring the spherical concave mirror radius of curvature with a laser rangefinder
-
Published:2023-04
Issue:2
Volume:47
Page:
-
ISSN:0134-2452
-
Container-title:Computer Optics
-
language:
-
Short-container-title:Computer Optics
Author:
Sakharov A.A., ,Zhivotovsky I.V.,Karasik V.E.,Patrikeeva A.A., , ,
Abstract
An instrument and technique for assessing errors of measuring optical surface radius of curvature with a laser rangefinder are presented. Errors of optical instrument alignment with a wave-front sensor are shown to influence the accuracy of measuring the mirror radius. Errors of the rangefinder-aided technique for measuring the surface radius are estimated. A computer analysis shows that the developed scheme of misalignment measurement allows a relative error of 0.02 – 0.3 % to be attained for mirrors ranging in radius from 1 m to 10 m. The choice of the accuracy characteristics of the rangefinders used for measuring the optical surface radius of curvature is justified.
Publisher
Samara National Research University
Subject
Electrical and Electronic Engineering,Computer Science Applications,Atomic and Molecular Physics, and Optics
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. A common points selection method based on the uniformity of FLTMMS for pre-alignment of particle accelerators;Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment;2024-05