Validation of the Sysmex XN analyser and Blood Bank mode for the quality and safety of donor blood and transfusion products

Author:

Valina Ondrej1,Vankova Ida2,Mintzas Konstantinos1,Knappova Klara2,Gasova Zdenka2

Affiliation:

1. Haematology Department Sysmex Europe SE Norderstedt Germany

2. Apheresis and Transfusion Departments Institute of Haematology and Blood Transfusion Prague Czech Republic

Abstract

AbstractObjectivesOur objective was to compare the measurement of residual white blood cell (rWBC) and residual red blood cell (rRBC) counts in blood products using the XN Blood Bank mode and the laboratory standard operating procedures for manual counts. In addition, to compare the whole blood complete blood count (CBC) values of blood donors and the quality of blood products using the Sysmex XN analyser versus the XS‐1000i analyser.Materials and MethodsFor blood donors, 190 samples from blood or apheresis donors were analysed on both the Sysmex XS‐1000i and XN‐1000 analysers and the mean values of six CBC parameters were compared: the white blood cell count (WBC), the red blood cell count (RBC), haemoglobin (HGB), haematocrit (HCT), the mean corpuscular volume (MCV), the platelet count (PLT). For blood products, 164 samples were collected: 13 Plasma products – whole blood, 9 Plasma products – apheresis, 36 RBC concentrates – whole blood, 30 PLT concentrates – buffy coats, 36 PLT concentrates – buffy coats – pooled and 55 PLT concentrates – apheresis.ResultsAll CBC parameters of the blood donors tested showed similar performance, with excellent correlation coefficients (r) ranging from 0.821 to 0.995. The majority of the blood products did not have a quantifiable number of residual cells, meaning the number of rWBC and rRBC, if present, was below the limit of quantitation (LoQ) of the different methods. rWBC were detected by Blood Bank mode in Plasma products – whole blood with a mean rWBC of 0.012 × 109/L and in PLT concentrates – buffy coats with a mean rWBC of 0.19 × 109/L. The correlation coefficient in both analysers for all three parameters (HGB, HCT, RBC) in RBC concentrates – whole blood was excellent, ranging from 0.95 to 0.99. For platelet count, r ranged from 0.98 to 0.99.ConclusionThe XN‐Series analyser, equipped with a Blood Bank mode, demonstrated reliable performance when used for blood donor evaluation, rWBC enumeration and measurement of end blood products.

Publisher

Wiley

Subject

Hematology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3