Affiliation:
1. Department of Anesthesiology Duke University Durham North Carolina USA
2. Department of Anesthesiology, Pain Division Duke University Durham North Carolina USA
Abstract
AbstractBackgroundOne of the unique advances in neuromodulation for chronic pain has been spinal cord stimulators (SCS) and dorsal root ganglion stimulators (DRG‐S). These devices have aided in conditions such as neuropathic pain, complex regional pain syndromes, failed back surgery, and peripheral neuropathies. With these benefits, however, complications from implantable stimulators have included lead fractures and migration. The authors reviewed a lead migration, kinking, and subsequent fracture event involving a patient with complex regional pain syndrome (CRPS) II, who was treated with a DRG‐S.Case PresentationThe case report follows this patient, from their past medical history to assessment of appropriate qualifications for neuromodulation, to successful surgical placement, to follow‐up care. The authors further monitored assessment of inefficacy of pain relief, and identification of lead migration and kinking through imaging. In the process of removal, due to lead stress, lead fracturing occurred. After lead removal, the leads were fully replaced, and the patient was followed up and experienced improved pain relief.ConclusionThe case report assesses probable mechanisms of lead fracture and considerations for physicians for future assessment and triage of neuromodulation efficacy.
Subject
Anesthesiology and Pain Medicine
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. Meloxicam/methocarbamol;Reactions Weekly;2024-06-22