Effect of implant shape and length on the accuracy of robot‐assisted immediate implant surgery: An in vitro study

Author:

Wang Yunxiao12ORCID,Yu Shimin12ORCID,Wang Yulan12ORCID,Feng Yi2ORCID,Yan Qi12ORCID,Zhang Yufeng123ORCID

Affiliation:

1. State Key Laboratory of Oral and Maxillofacial Reconstruction and Regeneration, Key Laboratory of Oral Biomedicine Ministry of Education, Hubei Key Laboratory of Stomatology, School and Hospital of Stomatology Wuhan University Wuhan China

2. Department of Oral Implantology, School and Hospital of Stomatology Wuhan University Wuhan China

3. Oral Biomaterials and Application Technology Engineering Research Center of Hubei Province Wuhan China

Abstract

AbstractObjectivesTo compare the accuracy of immediate implant placement of cylindrical implants (CI) and tapered implants (TI) of different lengths using a robotic dental implant system.Materials and MethodsCI and TI of three lengths (8, 10, and 12 mm) each were digitally planned and placed in a three‐dimensional printed extraction socket model under robotic guidance. There were six groups with three samples in each group, resulting in a total of 18 samples. Implant angular deviation, platform point deviation (total, lateral, depth), and implant apical point deviation (total, lateral, depth) were recorded and compared between the different groups.ResultsThe angular deviations for CI 8 mm, CI 10 mm, CI 12 mm, TI 8 mm, TI 10 mm, and TI 12 mm were 1.32° ± 0.19°, 1.03° ± 0.56°, 1.31° ± 0.38°, 1.27° ± 0.64°, 1.10° ± 0.43° and 1.05° ± 0.45°, respectively. The total deviations of platform and apical points for CI 8 mm, CI 10 mm, CI 12 mm, TI 8 mm, TI 10 mm, and TI 12 mm were 0.79 ± 0.18 mm, 0.77 ± 0.33 mm; 0.64 ± 0.21 mm, 0.55 ± 0.17 mm; 0.64 ± 0.37 mm, 0.65 ± 0.34 mm; 0.68 ± 0.26 mm, 0.71 ± 0.20 mm; 0.70 ± 0.12 mm, 0.66 ± 0.23 mm; and 0.71 ± 0.15 mm, 0.77 ± 0.29 mm, respectively, and had no significant differences.ConclusionsWithin the limitation of this study, acceptable accuracy can be achieved for both TI and CI using robotic systems. Our study demonstrated that the implant shape and length did not affect the accuracy of immediate implant placement under robotic guidance in vitro. However, further trials are required to confirm their efficacy in clinical practice.

Funder

Fundamental Research Funds for the Central Universities

Publisher

Wiley

Subject

Oral Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3