Diaphragmatic Myopotential Oversensing Caused by Change in Implantable Cardioverter Defibrillator Sensing Bandpass Filter

Author:

PLOUX SYLVAIN1,SWERDLOW CHARLES D.2,ESCHALIER ROMAIN3,MONTEIL BENJAMIN1,OUALI SANA4,HAÏSSAGUERRE MICHEL1,BORDACHAR PIERRE1

Affiliation:

1. Hôpital Cardiologique du Haut-Lévêque, CHU Bordeaux; Université Bordeaux, IHU LIRYC; Bordeaux France

2. Cedars-Sinai Heart Institute; Cedars-Sinai Medical Center; Los Angeles California

3. Image Science for Interventional Techniques (ISIT), UMR6284, and CHU Clermont-Ferrand, Cardiology Department, Clermont Université, Cardio Vascular Interventional Therapy and Imaging (CaVITI); Université d'Auvergne; Clermont-Ferrand France

4. Département de Cardiologie; Hôpital La Rabta; Tunis Tunisia

Funder

French Government, Agence Nationale de la Recherche au titre du programme Investissements d'Avenir

Publisher

Wiley

Subject

Cardiology and Cardiovascular Medicine,General Medicine

Reference12 articles.

1. Myopotential sensing by a dual chamber implantable cardioverter defibrillator: Two case reports;Deshmukh;J Cardiovasc Electrophysiol,1998

2. Inappropriate shocks from implanted cardioverter defibrillators caused by sensing of diaphragmatic myopotentials;Peters;J Interv Card Electrophysiol Int J Arrhythm Pacing,1998

3. Implantable cardiac defibrillator lead failure or myopotential oversensing? An approach to the diagnosis of noise on lead electrograms;Kowalski;Eur Eur Pacing Arrhythm Card Electrophysiol J Work Groups Card Pacing Arrhythm Card Cell Electrophysiol Eur Soc Cardiol,2008

4. Preventing overdiagnosis of implantable cardioverter-defibrillator lead fractures using device diagnostics;Swerdlow;J Am Coll Cardiol,2011

5. Diaphragmatic myopotential oversensing in pacemaker-dependent patients with CRT-D devices;Santos;Europace,2008

Cited by 9 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Low-frequency attenuation filter as a reversible cause of inappropriate shocks due to myopotential oversensing;HeartRhythm Case Reports;2021-09

2. Troubleshooting;Cardiac Pacing, Defibrillation and Resynchronization;2021-05-19

3. Inappropriate pacing due to T-wave oversensing;Journal of Thoracic Disease;2020-05

4. High incidence of diaphragmatic myopotential oversensing by a specific implantable cardioverter defibrillator;Pacing and Clinical Electrophysiology;2019-12-30

5. A review of specialized and automated features in implantable cardiac devices;Journal of Cardiovascular Electrophysiology;2019-11-21