Abstract
AbstractPurposeTo determine whether correcting the axial length (AL) measurement error of the IOLMaster 700 could improve the refractive prediction accuracy in silicone oil‐filled eyes.MethodsThis study included 265 cataract patients (265 eyes) with silicone oil tamponade who were scheduled for phacoemulsification with intraocular lens (IOL) implantation. The performances of various formulas, including Barrett Universal II, Emmetropia Verifying Optical, Hoffer‐QST, Kane, Ladas Super Formula, Pearl‐DGS, Radial Basis Function and traditional formulas (Haigis, Hoffer Q, Holladay 1 and SRK/T), were evaluated. The refractive prediction errors (PE) calculated with measured AL (ALmeas) and corrected AL with silicone oil adjustment (SOAL) were compared. Subgroup analysis was performed based on the ALmeas (<23 mm; 23–26 mm; ≥26 mm).ResultsUsing SOAL significantly reduced the hyperopic PE of formulas when compared to ALmeas (−0.05 to 0.17 D vs 0.15 to 0.38 D, p < 0.001). After applying AL correction, all formulas showed a lower mean absolute PE (0.47–0.57 D vs 0.50–0.69 D). The percentage of eyes within ±1.0 D of PE increased from 84.91%–88.68% to 89.81%–91.32% for new formulas and from 78.11%–83.40% to 85.66%–88.68% for traditional formulas, with the use of SOAL. Subgroup analysis showed that the majority of formulas with SOAL in prediction accuracy for eyes with an AL ≥26 mm (p < 0.05).ConclusionsThe refractive prediction accuracy in silicone oil‐filled eyes was improved by correcting the AL measurement error of the IOLMaster 700, especially for long eyes.
Funder
National Natural Science Foundation of China
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献