Contact force and catheter stability are predictive metrics of successful pulmonary vein isolation with high‐power short duration ablation in atrial fibrillation

Author:

Honarbakhsh Shohreh12ORCID,Martin Claire A.3,Dhillon Gurpreet4,Gallagher Mark4,Sohal Manav4,Chow Anthony1,Ahsan Syed1,Abbass Hakam1,Whittaker‐Axon Sarah1,Lambiase Pier D.1,Davies Mark J.5,Ginks Matthew5,Hunter Ross J.1

Affiliation:

1. The Barts Heart Centre, St Bartholomew's Hospital Barts Health NHS Trust London UK

2. William Harvey Research Institute Queen Mary's University of London London UK

3. Royal Papworth Hospital Cambridge UK

4. St George's University Hospitals NHS Foundation Trust London UK

5. Oxford University Hospitals NHS Foundation Trust Oxford UK

Abstract

AbstractIntroductionPreliminary data suggest that high power short duration (HPSD) ablation for pulmonary vein isolation (PVI) are safe. Limited data are available on its effectiveness. Aim was to evaluate HPSD ablation in atrial fibrillation ablation using a novel Qdot Micro catheter.Methods and ResultsProspective multicenter study evaluating safety and efficacy of PVI with HPSD ablation. First pass isolation (FPI) and sustained PVI was assessed. If FPI was not achieved additional ablation index (AI)‐guided ablation with 45 W was performed and metrics predictive of this were determined. Sixty‐five patients and 260 veins were treated. Procedural and LA dwell time was 93.9 ± 30.4 and 60.5 ± 23.1 min, respectively. FPI was achieved in 47 (72.3%) patients and 231 veins (88.8%) with an ablation duration of 4.6 ± 1.0 min. Twenty‐nine veins required additional AI‐guided ablation to achieve initial PVI with 24 anatomical sites ablated with the right posterior carina being the most common site (37.5%). A contact force of ≥8 g (area under the curve [AUC]: 0.81; p < 0.001) and catheter position variation of ≤1.2 mm (AUC: 0.79; p < 0.001) with HPSD were strongly predictive of not requiring additional AI‐guided ablation. Out of the 260 veins, only 5 (1.9%) veins showed acute reconnection. HPSD ablation was associated with shorter procedure times (93.9 vs. 159.4 min; p < 0.001), ablation times (6.1 vs. 27.7 min; p < 0.001), and lower rates of PV reconnection (9.2% vs. 30.8%; p = 0.004) compared to moderate power cohort.ConclusionsHPSD ablation is an effective ablation modality which results in effective PVI whilst maintaining a safety profile. Its superiority needs to be evaluated in randomized controlled trials.

Publisher

Wiley

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献