Red light instruments for myopia exceed safety limits

Author:

Ostrin Lisa A.1ORCID,Schill Alexander W.1

Affiliation:

1. University of Houston College of Optometry Houston Texas USA

Abstract

AbstractPurposeLow‐level red light (LLRL) therapy has recently emerged as a myopia treatment in children, with several studies reporting significant reduction in axial elongation and myopia progression. The goal of this study was to characterise the output and determine the thermal and photochemical maximum permissible exposure (MPE) of LLRL devices for myopia control.MethodsTwo LLRL devices, a Sky‐n1201a and a Future Vision, were examined. Optical power measurements were made using an integrating sphere radiometer through a 7‐mm diameter aperture, in accordance with ANSI Z136.1‐2014, sections 3.2.3–3.2.4. Retinal spot sizes of the devices were obtained using a model eye and high‐resolution beam profiler. Corneal irradiance, retinal irradiance and MPE were calculated for an eye positioned at the oculars of each device.ResultsBoth devices were confirmed to be Class 1 laser products. Findings showed that the Sky‐n1201a delivers laser light as a point source with a 654‐nm wavelength, 0.2 mW power (Ø 7 mm aperture, 10‐cm distance), 1.17 mW/cm2 corneal irradiance and 7.2 W/cm2 retinal irradiance (Ø 2 mm pupil). The MPE for photochemical damage is 0.55–7.0 s for 2–7 mm pupils and for thermal damage is 0.41–10 s for 4.25–7 mm pupils. Future Vision delivers the laser as an extended source subtending 0.75 × 0.325°. It has a 652‐nm wavelength, 0.06 mW power (Ø 7 mm aperture, 10 cm distance), 0.624 mW/cm2 corneal irradiance and 0.08 W/cm2 retinal irradiance (Ø 2 mm pupil). MPE for photochemical damage is 50–625 s for 2–7 mm pupils.DiscussionFor both of the LLRL devices evaluated here, 3 min of continuous viewing approached or surpassed the MPE, putting the retina at risk of photochemical and thermal damage. Clinicians should be cautious with the use of LLRL therapy for myopia in children until safety standards can be confirmed.

Funder

National Eye Institute

Publisher

Wiley

Cited by 12 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3