Irradiance from 12 LED light curing units measured using 5 brands of dental radiometers

Author:

Maucoski Cristiane12ORCID,Price Richard Bengt2ORCID,Arrais Cesar Augusto Galvão1ORCID

Affiliation:

1. Department of Restorative Dentistry State University of Ponta Grossa Ponta Grossa Brazil

2. Department of Dental Clinical Sciences Dalhousie University Halifax Nova Scotia Canada

Abstract

AbstractObjectiveTo evaluate the accuracy of five brands of radiometers in reporting the irradiance (mW/cm2) from twelve brands of LCUs compared to a ‘Gold Standard’ (GS) reference obtained from a hand‐held laboratory‐grade radiometer.Materials and MethodsThe irradiance was measured from two examples of twelve brands of previously used LCUs on two examples of five brands of dental radiometers. The emission spectrum was also obtained. Irradiance data from each brand of LCU against each meter was analyzed using the Shapiro–Wilk test for normality. The irradiance values were subjected to a two‐way ANOVA followed by Bonferroni tests for each LCU brand. Finally, a descriptive analysis was made using a 95% confidence interval around the mean irradiance.ResultsThe power output from the LCUs ranged from 271 mW to 1005 mW. Among the tested radiometers, only the Bluephase Meter II could accurately report the irradiance from 11 out of the 12 brands of LCU evaluated in this study. When measured using the “GS” system, the mean irradiance values from the two examples of nine brands of previously used LCU were not always within ±10% of the irradiance values stated by the manufacturer.ConclusionsThe mean irradiance values from 9 of the 12 brands of used LCUs were beyond ±10% of the irradiance values stated by the manufacturer. Only the Bluephase Meter II could accurately report the irradiance from 11 out of the 12 brands of LCU evaluated in this study.Clinical SignificanceThere was a wide range in the power output from the LCUs tested. It was impossible to accurately measure the irradiance from all the LCUs using the dental radiometers examined. However, dental radiometers should still be used in dental offices to monitor the light output from LCUs and verify that they are working correctly before they are used on patients.

Funder

Coordenação de Aperfeiçoamento de Pessoal de Nível Superior

Dalhousie University

Mitacs

Publisher

Wiley

Subject

General Dentistry

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3