Analytical performance of feline plasma on current Heska and IDEXX point‐of‐care biochemistry analyzers compared with a commercial laboratory analyzer

Author:

Baral Randolph M.1ORCID,Jaensch Susan M.2ORCID,Hayward Douglas A.2ORCID,Freeman Kathleen P.3ORCID

Affiliation:

1. University of Sydney/Paddington Cat Hospital Sydney Australia

2. Vetnostics Sydney Australia

3. Veterinary Information Network Davis California USA

Abstract

AbstractBackgroundPoint‐of‐care (POC) biochemistry analyzers are widely used in small animal clinical practice but infrequently independently assessed for performance.ObjectiveTo assess the performance of two current model point‐of‐care biochemistry analyzers (Heska Element DC and IDEXX Catalyst) compared with a commercial laboratory analyzer (Cobas 8000).MethodsOne hundred twenty‐one cats from a feline hospital population were sampled with plasma results from a single lithium heparin tube assessed on all three analyzers. Plasma biochemistry results from each POC analyzer were compared with the commercial laboratory analyzer using Bland–Altman difference plots and by determining whether the limits of agreement (LOAs) (95% of differences) fell within various quality goals after correcting for inherent bias.ResultsOnly 7 of 14 analytes on the Heska analyzer and 2 analytes on the IDEXX analyzer attained the most stringent LOA quality goal, which was being within desirable total error based on biologic variation (TEdes). The number of analytes achieving quality goals increased with less stringent standards such as American Society of Veterinary Clinical Pathologists allowable total error (ASVCP TEA) guidelines or if <95% of clinical comparisons reaching these quality goals is considered acceptable. Widespread bias was found between both POC analyzers and the commercial laboratory analyzer.ConclusionsThe performance of both POC biochemistry analyzers was variable compared with a commercial laboratory analyzer. Performance goals were only able to be attained after the bias for each analyzer was accounted for by offsetting the LOAs and quality goals set by the mean bias for each analyte on each analyzer.

Publisher

Wiley

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3