Author:
Baines A. D.,Timanoff M.,Wong J.,Paul W.
Abstract
The construction of a simple device for continuous monitoring and recording of plasma indocyanine green concentration in small animals is described. The apparatus is designed to hold a vascular bed such as the ear or the omentum and to transilluminate it with white light from a fiber optic bundle. The transmitted light is divided by a dichroic mirror. Appropriate filters select a narrow band of frequencies from each segment. There are two photovoltaic cells, one sensitive to the dye and the other insensitive to it. Both are approximately equally reactive to changes in transilluminence caused by changes in blood content or hematocrit and are relatively insensitive to changes in oxygen saturation of the transilluminated blood. Reproducible estimates of cardiac output have been obtained with the injection of 300 μg indocyanine green in a volume of 20 μl with the densitometer applied to the small intestine or to the ear of the animal. The device responds linearly to increasing plasma indocyanine green concentrations and can be calibrated with less than 0.1 ml of blood.
Publisher
Canadian Science Publishing
Subject
Physiology (medical),Pharmacology,General Medicine,Physiology
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献