Abstract
The techniques and difficulties encountered in measuring temperatures to the highest precision with platinum resistance thermometers are discussed. It is shown that the relative drift of the resistance coils in the Mueller resistance bridge used for these measurements is less than a part per million per year. The intrinsic resistance of a platinum thermometer is comparatively unstable, and results showing some effects of cold work and heat treatment on several thermometers are given.As each precision temperature determination involves the resistance of the thermometer at the triple point of water, extensive measurements have been carried out to obtain information on: (a) the reproducibility of temperature in particular cells, (b) the variation in temperature among cells, and (c) the long term stability of cell temperatures.The limiting uncertainties in temperature measurements due to variation in the bridge, the thermometers, and the triple point cells are each of the order of 10−4 °C.