Author:
Rayner D. M.,McKinnon A. E.,Szabo A. G.,Hackett P. A.
Abstract
The photomultipliers used in single photon counting fluorescence lifetime Systems have a time response function which varies with wavelength. This variation results in a reduced confidence in the kinetic parameters obtained after deconvolution of the lamp intensity decay profile measured at the excitation wavelength from the sample emission decay profile measured at the emission wavelength. Moreover in the case of multi-component decays, inaccurate kinetic parameters are calculated because of this PMT variation. Using weighted residual plots, we demonstrate how one may make a satisfactory subjective appraisal of the adequacy of the deconvolution and the physical model of the decay processes. We have developed a ratio correction technique which accounts for the PMT time response variation with wavelength. This technique requires no apriori assumption of the nature of the PMT effect. Only when the data are corrected using the ratio correction technique can sufficient confidence in the accuracy of the singlet excited state parameters and the validity of the physical model be obtained.
Publisher
Canadian Science Publishing
Subject
Organic Chemistry,General Chemistry,Catalysis
Cited by
93 articles.
订阅此论文施引文献