Abstract
Objective: The purpose of the trial is to comparatively analyze an electronic, pressure-calibrated probe third generation Parometer (Orange) and a standard, manual measurement probe WHO 621 (C type) in the context of taking periodontal variables when assessing periodontal status in childhood. Materials and methods: The subject of the study were 28 children aged between 12 and 14 years (12 boys and 16 girls). All patients were clinically examined, and the data were recorded on a specially prepared card. The recorded clinical variables contain: Assessment of oral hygiene habits (type of toothbrush, frequency of brushing); OHI as per Green Vermillion; Registration of dental status; Depth of gingival sulcus (on all teeth) with both types of probes; BOP (bleeding on probing), percentage of bleeding units with both types of probes; Taking into account the complete time needed to take the findings and the sensation of pain experienced by a digital rank scale during probing. Results and conclusion: The average depth of gingival sulcus measured with a mechanical periodontal probe was 1.62 mm, and with an electronic one - 1.38 mm (p <0.05). Values for BOP with both types of probes showed an average of 0.30 ± 0.29, which is 30% of all bleeding units examined (p> 0.05). The time for recording the periodontal indices with both probes is, on average, 10 minutes. In both probes, the discomfort of about grade 4 was observed according to the ranking scale used to read sensitivity (p <0.05). There is a more pronounced sensitivity when using an electronic periodontal probe.
Publisher
Peytchinski Publishing Ltd.
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献