Automated and Manual Methods for the Determination of Polyacrylamide and Other Anionic Polymers

Author:

Allison J.D.1,Wimberly J.W.1,Ely T.L.1

Affiliation:

1. Conoco Inc.

Abstract

Automated and Manual Methods for the Determination of Polyacrylamide and Other Polyacrylamide and Other Anionic Polymers Summary. The concentration of polyacrylamides is determined by precipitation with Hyamine 1622 Reagent I and by measurement of the amount precipitation with Hyamine 1622 Reagent I and by measurement of the amount of light scattered by the resulting turbidity. The analysis can be automated as well as adapted for field trials. The effects of anionic surfactants, changes in polyacrylamide molecular weight, and salinity are discussed. Introduction In this age of EOR, the use of polyacrylamides in the oil field is increasing steadily. Polyacrylamides are used in the viscosification of drilling muds, polymer-augmented mobility control, and the crosslinking of polymers for profile modification. Implementation of these techniques profile modification. Implementation of these techniques requires fast and accurate determination of polyacrylamides in a wide range of aqueous solutions. The method presented meets these criteria by being simple, accurate, and free presented meets these criteria by being simple, accurate, and free from most interferences. The analysis of polyacrylamide in water has historically been based on one of three principles: hydrolysis, with detection of the resulting ammonia; the starch iodide method (redox titration); and turbidimetry. Of these, the first two suffer interferences from background nitrogen compounds and redox processes, respectively. We have applied Crummett and Hummel, S turbidimetric method to the analysis of polyacrylamides in oilfield brine and production core tests. Hyamine 1622 (diisobutylphenoxyethoxyethyldimethylbenzylammonium chloride, Fig. 1) precipitates anionic polyacrylamides very effectively, providing detection limits of 0.5 to 1.0 ppm polyacrylamide. The reaction is shown in Fig. 2. Sodium polyacrylamide. The reaction is shown in Fig. 2. Sodium citrate is used to buffer the pH at 8.2. The sodium citrate also acts as a masking agent to prevent the precipitation of anionic surfactants (a potential interference) so that they can be tolerated up to the 200-ppm level. Changes in salinity and the presence of biocides (formaldehyde and/or glutaraldehyde) have no effect on the analysis. The presence of iron does interfere with the method. presence of iron does interfere with the method. Manual Method The manual method is straightforward and simple. The procedure for analysis follows. procedure for analysis follows.Pipette the appropriate aliquot of sample to a 50-mL mixing cylinder.Dilute to 30 mL with deionized water.Add 10 mL of 5% sodium citrate and mix.Add 10 mL of 0.4% Hyamine 1622 reagent and mix.Wait for 30 minutes for turbidity to develop.Measure turbidity in a 5-cm [1.97-in.] cell at 500 nm [5,000 A] with any standard colorimeter. The concentration of polyacrylamide in the sample is determined by extrapolation from the turbidity on a previously constructed calibration curve, such as that shown previously constructed calibration curve, such as that shown in Fig. 3. Observe in Fig. 3 that the method is linear to 20 ppm polyacrylamide. For higher levels of polyacrylamide, it may be necessary to dilute the sample further. polyacrylamide, it may be necessary to dilute the sample further. Because of the ease of manipulation, this method has been adapted to a field test, The turbidity development over 1 hour is shown in Fig. 4. Evidently, the reaction does not occur instantaneously and 40 minutes is required to achieve maximum turbidity. There is an acceptable error of 0. 1 ppm/min when the absorbance is measured at 30 minutes. Because the concentrations measured could be time-related, it is advantageous to automate the procedures; this adaptation is described in the following section. Adaptation to AutoAnalyzer The use of an AutoAnalyzer-type instrument ensures that the measurement is recorded at consistent reaction times because the same pump speed is used for both standards and samples. This gives identical times for mixing the reagent with the standard or sample, allowing the reaction to occur in the delay coil and the material to be introduced to the detector cell. The AutoAnalyzer manifold was set up as shown in Fig. 5. The standards and samples, in 2-mL cups, were loaded into the sampler tray with the samples placed immediately after the standards. With the system at reagent equilibrium (adjusted to steady baseline), the samples were aspirated and mixed with citrate and air. Reagent was then added and the mixing continued through a single mixing coil, where the turbidity complex forms. SPERE P. 184

Publisher

Society of Petroleum Engineers (SPE)

Subject

Process Chemistry and Technology

Cited by 12 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3