Method for assessing the accuracy of source positioning during brachytherapy

Author:

Kazlouski D. I.1ORCID

Affiliation:

1. N.N. Alexandrov National Cancer Centre of Belarus, agro-town Lesnoy

Abstract

A method for assessing the accuracy of source positioning using a new phantom for quality control procedures of applicators in brachytherapy is presented. A description is given of a phantom for performing measurements as part of quality control procedures for brachytherapy applicators, which are placed in the phantom in the form of a rod at a distance of 2–4 cm from the location of the ionization chamber. The air kerma strength was measured and the distance from the source to the ionization chamber was calculated. The measured values of the distances were compared with the values obtained by measuring the X-ray images of the applicator with a source located in it or radiopaque markers. The parameters recorded using the developed phantom are in good agreement with the parameters obtained using the X-ray machine installed in the procedural room (p > 0.05). The use of the proposed phantom will allow not only determination of the accuracy of the radiation source position in the applicator channel, but also verification of the irradiation plans for various types of applicators, including the use of shielding blocks.

Publisher

Publishing House Belorusskaya Nauka

Subject

Pharmacology (medical)

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3