Safety and Tolerability of Implanted Subcutaneous Cardioverter-Defibrillator Systems

Author:

Vereshchagina A. V.1ORCID,Uskach T. M.2ORCID,Sapelnikov O. V.1ORCID,Amanatova V. A.1ORCID,Grishin I. R.1ORCID,Kulikov A. A.1ORCID,Kostin V. S.1ORCID,Akchurin R. S.1ORCID

Affiliation:

1. National Medical Research Center of Cardiology named after academician E.I. Chazov

2. National Medical Research Center of Cardiology named after academician E.I. Chazov; Russian Medical Academy of Continuous Professional Education

Abstract

Aim. To study the safety and tolerability of the subcutaneous implantable cardioverter defibrillator (S-ICD) after implantation.Material and methods. The results of 33 patients with implanted S-ICD 6 months follow-up. The criteria for inclusion in the observational study were: age over 18 years, indications for primary or secondary prevention of sudden cardiac death. The exclusion criteria were indications for implantation of transvenous ICD (patients with sustained monomorphic ventricular tachycardia, the need for anti-bradycardia or resynchronization therapy), as well as patients with a QRS complex of more than 130 msec. All patients underwent a standard preoperative examination (routine blood tests, chest X-ray, transthoracic echocardiography), quality-of-life questionnaires and transesophageal echocardiography. At follow-up, patients were examined after 6 months after implantation, the device was interrogated and a quality-of-life questionnaire was completed. All episodes of shock therapy and complications were documented.Results. Male patients predominated (84%), with a mean age of 57 [43;62] years. Left ventricular ejection fraction was 30% [26;34]. The mean QRS duration was 100 [94;108] msec. According to the of 24-hour Holter ECG monitoring, episodes of unstable VT were recorded in 42.4% of patients. The most common indications for S-ICD implantation were dilated (33%) and ischemic cardiomyopathy (42%). Primary prevention was indicated in 97% of patients. At the end of the implantation of the S-ICD, the patients underwent a defibrillation test and device configuration. In 63.6% of cases, during automatic tuning, the device selected the primary perception vector. In 27.2% of patients, optimal recognition of the subcutaneous signal was observed in the secondary vector, and in 9.2% of patients, the alternative vector was favorable. All patients underwent two-zone programming. The conditional shock zone was programmed at an average rate of 192 beats/min (range 180-210 beats/min) and the shock zone was programmed at an average rate of 222 beats/min (range 220-240 beats/min). Perioperative complications occurred in two patients. During the follow-up period, no shocks were recorded in 27 patients. Adequate shocks for 6 months were recorded in two patients. During 6 months of observation, one lethal outcome was noted due to complications of viral pneumonia. During the observation period, there were no rehospitalizations for cardiovascular diseases.Conclusion. The use of S-ICD, even in patients with structural myocardial disease who do not require antibradycardia pacing, is effective in preventing SCD. The number of inadequate discharges and the number of complications in clinical practice is comparable to the data of multicenter studies. S-ICD implantation was not accompanied by a decrease in quality of life. Careful selection of candidates, along with state-of-the-art device programming, is an important parameter for the selection and success of S-ICD application.

Publisher

Stolichnaya Izdatelskaya Kompaniyaizdat

Subject

Pharmacology (medical),Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3