Development and ex-vivo validation of 36G polyimide cannulas integrating a guiding miniaturized OCT probe for robotic assisted subretinal injections

Author:

Abid Alexandre12,Duval Renaud31,Boutopoulos Christos12ORCID

Affiliation:

1. University of Montreal

2. Maisonneuve-Rosemont Hospital Research Centre

3. Maisonneuve-Rosemont Hospital

Abstract

We introduced and validated a method to encase guiding optical coherence tomography (OCT) probes into clinically relevant 36G polyimide subretinal injection (SI) cannulas. Modified SI cannulas presented consistent flow capacity and tolerated the typical mechanical stress encountered in clinical use without significant loss of sensitivity. We also developed an approach that uses a micromanipulator, modified SI cannulas, and an intuitive graphical user interface to enable precise SI. We tested the system using ex-vivo porcine eyes and we found a high SI success ratio 95.0% (95% CI: 83.1–99.4). We also found that 75% of the injected volume ends up at the subretinal space. Finally, we showed that this approach can be applied to transform commercial 40G SI cannulas to guided cannulas. The modified cannulas and guiding approach can enable precise and reproducible SI of novel gene and cell therapies targeting retinal diseases.

Funder

Vision Health Research Network

Fonds de recherche en ophtalmologie de l’Université de Montréal

Fonds de Recherche du Québec - Santé

Publisher

Optica Publishing Group

Subject

Atomic and Molecular Physics, and Optics,Biotechnology

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3