Method to quickly estimate T1 value by suppressing spin exchange relaxation and magnetic field gradient relaxation in atomic sensors

Author:

Wang Xuelei12ORCID,Li Jianli123,Qu Chunyu123,Cheng Yu12,Zhang Junjie12

Affiliation:

1. Hefei national laboratory

2. Institute of Large-scale Scientific Facility and Centre for Zero Magnetic Field Science, Beihang University

3. National Institute of Extremely-Weak Magnetic Field Infrastructure

Abstract

The accuracy of atomic sensor is determined by the transverse nuclear spin relaxation time (T2). As the longitudinal nuclear spin relaxation time (T1) is the upper limit of T2, it becomes an important metric for evaluating the merits of alkali vapor cells. However, the conventional method for measuring T1 of pulse delay takes dozens of minutes, which will increase the effect of random errors during the testing process. In order to solve this problem, a method to quickly estimate the value of T1 is proposed by suppressing the influence of spin exchange relaxation and magnetic field gradient relaxation terms on T2. Experimental results indicate that the accuracy is maintained at above 95%, and the testing time is reduced to less than 5%.

Funder

Agricultural Science and Technology Innovation Program

National Science Fund for Distinguished Young Scholars

National Natural Science Foundation of China

Publisher

Optica Publishing Group

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3