Abstract
All-optical ultrasound (OpUS) is an imaging paradigm that uses light to
both generate and receive ultrasound, and has progressed from benchtop
to in vivo studies in recent years, demonstrating
promise for minimally invasive surgical applications. In this work, we
present a rapid pullback imaging catheter for side-viewing B-mode
ultrasound imaging within the upper gastrointestinal tract. The device
comprised an ultrasound transmitter configured to generate ultrasound
laterally from the catheter and a plano-concave microresonator for
ultrasound reception. This imaging probe was capable of generating
ultrasound pressures in excess of 1 MPa with corresponding −6
dB bandwidths > 20 MHz. This enabled imaging resolutions as low as
45 µm and 120 µm in the axial and lateral extent
respectively, with a corresponding signal-to-noise ratio (SNR) of 42
dB. To demonstrate the potential of the device for clinical imaging,
an ex vivo swine oesophagus was imaged using the
working channel of a mock endoscope for device delivery. The full
thickness of the oesophagus was resolved and several tissue layers
were present in the resulting ultrasound images. This work
demonstrates the promise for OpUS to provide rapid diagnostics and
guidance alongside conventional endoscopy.